I-LAB Specifications
2-Glove Workstation
Hide SpecificationsInternal dimensions | 1213 mm Wide x 764 mm Deep x 897 mm High |
External dimensions | 2005 mm Wide x 1034 mm Deep x 2024 mm High |
Internal finish | Type 304 stainless steel, #4 |
External surfaces | RAL 7035 |
Wall thickness | 0.105″ (2.6 mm) |
Feedthrough | 4 x DN 40 ISO-KF blank |
Internal power | 1 x electrical feedthrough |
Internal power | 1 x fuse protected 6-outlet power strip |
Shelving | 2 x Height adjustable 304 stainless steel #4 finish. 30.6″ (778 mm) Wide x 9.7″ (247 mm) Deep |
Plumbing | 304 stainless steel tubing |
Filters | 1 x inlet/1 x outlet HEPA (99.99% vs 0.3 micron MPPS) |
Lighting | External fluorescent light |
Power | 100-120 or 200-240 VAC 50/60 Hz 20 Amp |
Non-Modular Box Design
View SpecificationsStand with casters and leveling feet | Leveling feet adjustability |
Large Antechamber, 15″ (381 mm) Internal Diameter x 24″ (609 mm) Internal Length
View SpecificationsBolt | On left or right side |
Internal finish | 304 stainless steel #4 surface finish |
External finish | RAL 7035 |
Internal tray | 304 stainless steel #4 surface finish |
Tray size | 11.25″ (285 mm) Wide x 22.5″ (571 mm) Long |
Internal/external travel | 13.9″ (353 mm) |
Anodized aluminum doors, vertical opening with gas piston | |
Manual evacuate/refill valves | |
Analog vacuum gauge | 0 to -30″ Hg |
Leak rate | < 10-5 mbar l/s |
Small Antechamber, 6″ (152 mm) Internal Diameter x 17.3″ (441 mm) Internal Length
View SpecificationsInternal finish | 6061 anodized aluminum |
External finish | RAL 7035 |
Internal/external doors | 6061 anodized aluminum |
Manual evacuate/refill valve | |
Analog vacuum gauge | 0 to -30″ Hg |
Leak rate | < 10-5 mbar l/s |
Polycarbonate (Abrasion Resistant) Window 3/8″ (9.5 mm) Thick
View SpecificationsViewing size | 41.8″ (1061 mm) Wide x 30.2″ (767 mm) High |
Circular Glove Ports
View SpecificationsBlack Delrin, 8″ glove port | 0.015″ (0.4 mm) thick butyl rubber gloves, hand size 9.75 |
I-Lab Gas Management System
View SpecificationsSingle column gas purification system | |
O2 and H2O purity level | |
O2 capacity = | 43 L |
H2O capacity = | 1800 g |
Automated purifier column valves | Interlocked with the blower. Blower will not run if the valves fail to open |
Variable speed circulation blower | 120 cfm max flow rate blower |
Max flow rate through purifier column | 101m3/h (60 CFM) 60 cfm |
PLC Control System
View SpecificationsColor touch screen |
Automatic Pressure Control
View SpecificationsAutomatic pressure control | (+/- 15 mbar) |
Manual pressure control | Footswitch |
Fully Automatic Regeneration Sequence
View SpecificationsManual purge valve |
Vacuum Pump
View SpecificationsVacuum pump | 10 cfm (17 m3/h) rotary vane pump with oil mist eliminator & oil return kit |
Integrated Oxygen Analyzer
View SpecificationsRange 0-1000 ppm |
Zirconia sensor as standard |
Or electrochemical sensor for solvent compatibility |
Integrated Moisture Analyzer
View SpecificationsRange 0-1000 ppm |
Overall System Leak Rate
View Specifications<0.05vol%/h (ISO 10648-2) |
System Gas Requirements
View SpecificationsWorking gas | ³ 99.995% N2 or Ar |
Working gas pressure | 55 psi (3.8 bar) |
Working gas flow rate | 250 l/min |
Regeneration gas | 3-7% H2, balance N2 or Ar |
Regeneration gas pressure | 5-7 psi (0.3-0.5 bar) |
Regeneration gas flow rate | 30 l/min |