Argon-10 Gas Management System
24/7 operation of enclosures up to 12m3
Inert’s Argon-10 dual column Gas Management System is equipped with two high-capacity purification columns, 0-185 CFM variable speed blower and 3-inch diameter gas loop for increased flow rate of inert gas. This system is perfect for high throughput or large enclosures that require a < 1ppm O2 and H2O inert atmosphere.
| Regenerable Purification Coulmn | Dual Regenerable Purification Coulmn | Quad Regenerable Purification Column | Large Capacity Solvent Removal | Regenerable Large Capacity Solvent Removal | Max Enclosure Volume | |||||
| Argon-10 | 12m3 | |||||||||
|
||||||||||
Specifications
| Dual purifier column design allows 24/7 operation | Closed loop recirculation |
| Piping and fittings | 304 stainless steel |
| Circulation piping | 80 mm diameter with DN 80 ISO-KF flanges for easy connection to customer enclosure |
| External dimensions | 48″ (1219 mm) Wide x 50″ (1270 mm) Deep x 62″ (1575 mm) High |
| External surfaces | RAL 7035 |
| Internal Features | |
| Internal cooling fans | |
| Interior LED for maintenance | |
| Purity Level | |
| < 1ppm O2 and H2O Purity Level | |
| O2 capacity = | 217 L per purifier column |
| H2O capacity = | 9000 g per purifier column |
| Column Valves | |
| Automated purifier column valves | Interlocked with the blower. Blower will not run if the valves fail to open |
| Circulation Blower | |
| Variable speed circulation blower | 0 to 100% |
| Max flow rate = | 185 cfm (315 m3/h) |
| PLC Control | |
| 6″ color touch screen | |
| Trending screen | Pressure, O2 and H2O value is logged for 24 hours |
| Pressure Control | |
| Automatic pressure control | (+/- 15 mbar) |
| Manual pressure control | Footswitch |
| Regeneration Sequence | |
| Automatic blower restart after regeneration | |
| Purge | |
| Purge valve | O2 level, timer or manually controlled |
| Vacuum | |
| Vacuum pump | 22 cfm (38 m3/h) rotary vane pump with oil mist eliminator & oil return kit |
| Eco Mode Features | |
| Vacuum pump timer | Energy reduction shuts pump off |
| Automated blower speed control | O2 level adjusts blower speed |
| Component Cycle Data | |
| Component cycle data | Vacuum pump, column usage, blower usage |
| System Gas Requirements | |
| Working gas | 3 99.995% Ar |
| Working gas pressure | 55 psi (3.8 bar) |
| working gas flow rate | 250 l/min |
| Regeneration gas | 3-7% H2, balance Ar |
| Regeneration gas pressure | 5-7 psi (0.3-0.5 bar) |
| Regeneration gas flow rate | 30 l/min |