I-Lab 2GB
Glovebox System
Overview
The I-Lab line of Glovebox and Gas Management Systems are designed as an economical solution for inert research applications.
The I-Lab Gas Management Systems provide a < 1ppm O2 and H2O inert atmosphere required for a multitude of air-sensitive applications.
I-LAB Specifications
Internal dimensions | 1213 mm Wide x 764 mm Deep x 897 mm High |
External dimensions | 2005 mm Wide x 1034 mm Deep x 2024 mm High |
Internal finish | Type 304 stainless steel, #4 |
External surfaces | RAL 9011 (Graphite Black) |
Wall thickness | 0.105″ (2.6 mm) |
Feedthrough | 4 x DN 40 ISO-KF blank |
Internal power | 1 x electrical feedthrough |
Internal power | 1 x fuse protected 6-outlet power strip |
Shelving | 2 x Height adjustable 304 stainless steel #4 finish. 30.6″ (778 mm) Wide x 9.7″ (247 mm) Deep |
Plumbing | 304 stainless steel tubing |
Filters | 1 x inlet/1 x outlet HEPA (99.99% vs 0.3 micron MPPS) |
Lighting | External LED light |
Power | 100-120 or 200-240 VAC 50/60 Hz 20 Amp |
Large Antechamber, 15″ (381 mm) Internal Diameter x 24″ (609 mm) Internal Length
View Specifications
Bolt | On left or right side |
Internal finish | 304 stainless steel #4 surface finish |
External finish | RAL 9011 (Graphite Black) |
Internal tray | 304 stainless steel #4 surface finish |
Tray size | 11.25″ (285 mm) Wide x 22.5″ (571 mm) Long |
Internal/external travel | 13.9″ (353 mm) |
Anodized aluminum doors, vertical opening with gas piston | |
Manual evacuate/refill valves | |
Analog vacuum gauge | 0 to -30″ Hg |
Leak rate | < 10-5 mbar l/s |
Small Antechamber, 6″ (152 mm) Internal Diameter x 17.3″ (441 mm) Internal Length
View Specifications
Internal finish | 6061 anodized aluminum |
External finish | RAL 9011 (Graphite Black) |
Internal/external doors | 6061 anodized aluminum |
Manual evacuate/refill valve | |
Analog vacuum gauge | 0 to -30″ Hg |
Leak rate | < 10-5 mbar l/s |
Single column gas purification system | |
O2 and H2O purity level | |
O2 capacity = | 43 L |
H2O capacity = | 1800 g |
Automated purifier column valves | Interlocked with the blower. Blower will not run if the valves fail to open |
Variable speed circulation blower | 120 cfm max flow rate blower |
Max flow rate through purifier column | 101m3/h (60 CFM) 60 cfm |