Argon-10 Gas Management System
24/7 operation of enclosures up to 12m3
Overview
Inert’s Argon-10 dual column Gas Management System is equipped with two high-capacity purification columns, 0-185 CFM variable speed blower and 3-inch diameter gas loop for increased flow rate of inert gas. This system is perfect for high throughput or large enclosures that require a < 1ppm O2 and H2O inert atmosphere.
Regenerable Purification Coulmn | Dual Regenerable Purification Coulmn | Quad Regenerable Purification Column | Large Capacity Solvent Removal | Regenerable Large Capacity Solvent Removal | Max Enclosure Volume | |||||||||||
Argon-10 | 12m3 | |||||||||||||||
|
Specifications
Dual purifier column design allows 24/7 operation | Closed loop recirculation |
Piping and fittings | 304 stainless steel |
Circulation piping | 80 mm diameter with DN 80 ISO-KF flanges for easy connection to customer enclosure |
External dimensions | 48″ (1219 mm) Wide x 50″ (1270 mm) Deep x 62″ (1575 mm) High |
External surfaces | RAL 7035 |
Internal Features | |
Internal cooling fans | |
Interior LED for maintenance | |
Purity Level | |
< 1ppm O2 and H2O Purity Level | |
O2 capacity = | 217 L per purifier column |
H2O capacity = | 9000 g per purifier column |
Column Valves | |
Automated purifier column valves | Interlocked with the blower. Blower will not run if the valves fail to open |
Circulation Blower | |
Variable speed circulation blower | 0 to 100% |
Max flow rate = | 185 cfm (315 m3/h) |
PLC Control | |
6″ color touch screen | |
Trending screen | Pressure, O2 and H2O value is logged for 24 hours |
Pressure Control | |
Automatic pressure control | (+/- 15 mbar) |
Manual pressure control | Footswitch |
Regeneration Sequence | |
Automatic blower restart after regeneration | |
Purge | |
Purge valve | O2 level, timer or manually controlled |
Vacuum | |
Vacuum pump | 22 cfm (38 m3/h) rotary vane pump with oil mist eliminator & oil return kit |
Eco Mode Features | |
Vacuum pump timer | Energy reduction shuts pump off |
Automated blower speed control | O2 level adjusts blower speed |
Component Cycle Data | |
Component cycle data | Vacuum pump, column usage, blower usage |
System Gas Requirements | |
Working gas | 3 99.995% Ar |
Working gas pressure | 55 psi (3.8 bar) |
working gas flow rate | 250 l/min |
Regeneration gas | 3-7% H2, balance Ar |
Regeneration gas pressure | 5-7 psi (0.3-0.5 bar) |
Regeneration gas flow rate | 30 l/min |